July 23, 2026 · Resuscitation · DOI: 10.1016/j.resuscitation.2026.111217

Measurement of tidal volumes and respiratory rate during regular and asynchronous intra-arrest ventilation using the EOlife X device: A method-comparison study in a porcine cardiac arrest model

Listen to this summary

The authors aimed to evaluate the accuracy of the EOlife X ventilation feedback device during both regular and intra-arrest ventilation in a porcine cardiac arrest model. They found that while the device performed within clinically acceptable limits during regular ventilation, it significantly deviated from these limits during intra-arrest ventilation, likely due to the effects of chest compressions on airflow measurements. This study highlights the need for specific validation of ventilatory monitoring devices under CPR conditions.

Simon Orlob, David Purkarthofer, Wolfgang J Kern, Benjamin Hackl, Jonas Dunz, Stefanie Eder, Frederik Bartz, Sonja Fruhwald, Jan Wnent, Jan-Thorsten Gräsner, Martin Holler, Judith Martini, Gabriel Putzer