November 23, 2026 · Heart rhythm · DOI: 10.1016/j.hrthm.2026.11.028

Implantable Cardiac Monitor for Diagnosis and Continuous Monitoring of Type 1 Brugada Electrocardiographic Pattern.

Listen to this summary

This study asked whether implantable cardiac monitors (ICMs) with remote transmission can detect spontaneous type 1 Brugada ECG patterns in asymptomatic patients who showed the pattern only during ajmaline challenge. In 59 patients, with the device placed at a left upper thoracic position identified by surface potential mapping, the pattern was documented in 49% over a median of about 21.6 months, most often within two months of implantation. The authors conclude that continuous long-sensing-vector ICM monitoring is feasible for detecting spontaneous BrECG and reveals substantial temporal variability in its expression, with detection more frequent at night and at lower heart rates.