November 10, 2026 · Journal of cataract and refractive surgery · DOI: 10.1097/j.jcrs.0000000000002104

Effects of Ultrasound Modality, Viscosurgical Devices, and Automated Energy Modulation on Intraocular Thermal Exposure During Phacoemulsification.

Listen to this summary

This ex vivo porcine study asks what factors drive intraocular heat generation during phacoemulsification and whether automated energy modulation (Thermal Sentry) can limit it. Across 80 eyes, longitudinal ultrasound produced the greatest temperature rise, 4D mode the least, and OVD use increased heating in 4D mode, with temperature closely tracking delivered energy and cumulative dissipated energy. Thermal Sentry lowered anterior chamber temperature across all conditions and reduced incision temperature for torsional and longitudinal modes, suggesting automated energy modulation is an effective way to limit thermal exposure.