November 18, 2026 · Journal of cataract and refractive surgery · DOI: 10.1097/j.jcrs.0000000000002103

Optimal target offset under asymmetric refractive loss and discrete intraocular lens power steps.

Listen to this summary

This analysis asks what refractive target minimizes expected loss when hyperopic and myopic outcomes carry unequal penalties and intraocular lens powers come in discrete steps. Under Gaussian prediction error and piecewise-linear loss, a conventional -0.25 D target with a 0.40 D prediction standard deviation implies that hyperopia is weighted about 2.8 times as heavily as myopia. Sensitivity analyses that penalize myopia more heavily shift the hypothetical optimal target to between +0.17 and +0.39 D, and the author concludes that selecting the nearest available lens power adds little expected loss when lens steps are no larger than the prediction standard deviation.