September 23, 2026 · Regional anesthesia and pain medicine · DOI: 10.1136/rapm-2026-108277

Asymptomatic intrathecal baclofen pump catheter failure discovered during routine reservoir replacement: a case report.

Listen to this summary

This case report describes a 30-something woman with cerebral palsy who had intrathecal baclofen therapy since 2011 and reported no functional decline, despite sluggish CSF flow noted in 2019. During a 2025 pump reservoir replacement, the catheter was found completely fractured at its fascial insertion point, requiring partial laminectomy and dural repair, after which her symptoms improved with new hardware and retitration. The authors conclude that subjective clinical benefit is an unreliable indicator of catheter patency, and that aging ITB systems warrant proactive radiographic monitoring and timely surgical revision.